DAISI MT Interferometer


Just like for the single fiber system the software for DAISI-MT was designed from ground up with many automation features.
Non-compressed, real time and high quality images are transferred from the hardware to the software via a USB2.0 high speed link at speeds up to 100 images per second in addition to the automation and control commands.
DAISI-MT is portable and can be interfaced to laptop or desktop computers.
All calibration steps are automated and embedded into a user-friendly software interface in order to yield error-free and reliable measurements.

Download DAISI-MT product datasheet



MECHANICS & ELECTRONICS


...White-Light interferometers' mechanical and electronic design is more challenging than Red-Light (monochromatic) interferometer's design. The whole optical system must be translated over several tens of microns with nanometer precision and in a perfectly linear fashion.
The DAISI-MT system does just that: based on a 30 microns range closed-loop piezoelectric transducer, the surface of the sample under test is scanned and mapped smoothly, precisely and quickly  in less than 10 seconds.

Key Features :
  Single unit for measurement of most types of Multi Fiber ferrules (MT-X, MPO, Mini-MT (MT-RJ), etc.) PC and APC.
  Measure up to 100 fibers in a single scan with DAISI-MT mega-pixel high resolution camera.
  White-Light interferometer.
  Vibration insensitive. No specific isolation table required.
  Closed-loop high precision 30 microns Z scanning range.
  Non-contact measurement.
  Rapid autofocus.
  One button easy operation.
  No exterior moving parts or adjustment screw
  Easy and fast switching from PC to APC, no change of ferrule holder required.
  Fast and automated measurement of endface parameters as required by standards.
  Accurate and repeatable measurements.
  Can be interfaced to laptop computers, only one USB2.0 link required.
  Compliance with Industry Standards for Interferometer Measurements
  High resolution 2D & 3D surface profiles
  Measurement Report and History Report in Excel
 

    

  MPO-12 APC ferrule measurement example.

                                    


                                MPO-12 APC endface measurement example.
 
 

 
Parameter *Repeatability/Reproducibility Range
Radius (mm) 0.5% / 0.5% 3 to flat
X & Y angles (°) 0.002 / 0.01 ±1° deviation from 0°(PC) or 8°(APC)
Fibre Height (microns) 0.01 / 0.01 up to 20
Measurement Speed (sec. for a 12 fibers ferrule)   13
AutoFocus speed (sec.)   10
Field of View (mm)   configurable
3.2 x 2.5 max.
Lateral Resolution (microns)   configurable
2.5 max.
Wavelength (nm)   White LED
Power requirements 12V 25VA

* 1 Sigma values based on the measurement of an MT-RJ connector with 2.5µm high fibers and 95% valid pixels.
Repeatability values calculated from 50 consecutive measurements without interaction
on connector between measurements.
Reproducibility values calculated from 50 consecutive measurements while removing and inserting
connector in ferrule holder between measurements

 

        

 